Operation Multi 3630/3620 IDS
18 ba77045e01 09/2019
4.3.2 Measurement settings of the sample
The measurement settings for the sample are defined when the sample is reg-
istered within a sample.
Default settings are printed in bold.
The general settings for the set apply to the entire meter. All other
settings always apply to all OxiTop
®
-IDS measuring heads in the
current set.
Menu item Possible setting Explanation
Sample name max. 8 characters The 8 characters can be freely
selected
Preselected is an A.
An index is also assigned autimati-
cally.
Measuring range
(for BOD standard)
0 ... 40 / 80 / 200 mg/l
0 ... 400 mg/l
0 ... 800 / 2000 / 4000 mg/l
The measuring range expected for the
sample can be selected from 7 perma-
nent ranges.
Selecting the measuring range defines
the Sample volume. The correspond-
ing Sample volume is displayed as
Info.
Calculation
(for BOD special)
Measuring range
Sample volume
Bottle volume
Parameter to be calculated
The parameters not calculated are set
in the Settings menu.
Settings The maximum setting ranges quoted here can be restricted by other set-
tings.
The setting selected in the Calculation menu is not available here.
Measuring range 0 ... 400000 mg/l Measuring range
Sample volume 10 ... [Bottle volume - 10] ml Sample volume
Bottle volume 50 ... 9999 ml Bottle volume
Dilution 1 + 0 ... 99 Dilution of the sample
(1 part sample + x parts water)