What to do if... Multi 3630/3620 IDS
32 ba77045e01 09/2019
7 What to do if...
Cause Remedy
An OxiTop
®
-IDS mea-
suring head cannot be
registered for a sam-
ple
The selected OxiTop
®
-IDS measur-
ing head is already registered for a
sample
Example: A reset was carried out at
an OxiTop
®
-IDS measuring head
registered for a sample
Register a free OxiTop
®
-IDS mea-
suring head
Release the OxiTop
®
-IDS measur-
ing head
Delete the individual OxiTop
®
-
IDS measuring head from the
sample for which it is registered
Delete the set in which the
OxiTop
®
-IDS measuring head is
registered.
No measurement
curves are displayed
although the measure-
ment has already been
running for an
extended period of
time
The AutoTemp phase is still run-
ning
The temperature control is incor-
rectly set or defective
An OxiTop
®
-IDS measuring head
cannot be addressed
Wait for the end of the AutoTemp
phase then start the evaluation
Temper the sample more precisely
See error symptom,
A OxiTop
®
-IDS measuring head
cannot be addressed
An measuring head
cannot be addressed
The radio link was switched off at
the OxiTop
®
-IDS measuring
head, e.g. to save energy
The battery of the OxiTop
®
-IDS
measuring head is empty
The radio link between OxiTop
®
-
IDS measuring head and meter is
interrupted
The OxiTop
®
-IDS measuring
head is defective
If necessary, read any measure-
ment data available from the
OxiTop
®
-IDS measuring head
Exchange the battery (all measure-
ment data are lost)
Remove any barriers and sources
of interference from the radio link
(see operating manual of the WLM
system).
Re-establish the connection
Exchange the defective OxiTop
®
-
IDS measuring head
Contact the service department
The Multi 3630/
3620 IDS meter does
not react
The batteries of the meter are
empty
Operating condition undefined or
EMC load unallowed
The meter is defective
Insert new batteries or connect the
power pack
Carry out a reset at the meter
(simultaneously press the keys
<MENU/ENTER> and <On/
Off>)
Contact the service department