Temperature control accuracy
*1
±1℃ (Chamber center, 210℃, Damper: fully closed)
Temperature distribution
accuracy *1
±2.5℃ (Set temp.: 210℃, Damper: full opened)
±2.8℃ (Set temp.: 260℃, Damper: full closed)
(Set temp.: 210℃,
Damper: full closed)
(Room temp. to max temp) Approx. 80min.
(210℃~50℃) Approx. 150min
Stainless SUS tube heater
Inner dia. 30mm x 2 (top panel)
Inner dia. 30mm x 2 (rear panel)
Temperature control
system
PID control via microcomputer
Digital setting by menu key and up/down keys
Temperature reading display: green 4-digit digital LED
Temperature setting display: orange 4-digit digital LED
1min~99hrs, 59min or 100hrs~999.5hrs (wait function included), res. 1 or 10min
Fixed temperature, Quick Auto Stop, Auto Stop, Auto Start, Program
3 patterns, 30 steps in program mode, repeat function included
Lock function, Auto recovering after power failure, Calibration offset
Self-diagnostic
functions
Failure of Sensor, heater, SSR, memory, internal communication,
temperature inputting circuit, automatic overheating prevention device,
overheating prevention device, measurement temperature
Main Circuit breaker, Overheating prevention device
External dimensions
(W×D×H mm) *1
Internal dimensions
(W×D×H mm)
Chamber rack load capacity
Number of tiers/rack support
pitch