13
Installation
Environmental Requirements
Provide ample space surrounding the sterilizer
Provide ample space surrounding the unit as
indicated below.
・Space for lid to open...13 inches or more
・Space to open door.....10.2 inches or more
・Space for ventilation...5.9 inches or more
Do not install under a low ceiling
An abundant amount of steam will spout out of the lid
just after operation. Do not install the sterilizer under
any electrical components such as an alarm.
Lock casters
Be sure to fasten the two front casters
after positioning the unit.
Release
Lock