EasyManua.ls Logo

Yamato SM310 - STERILIZE Procedure Steps

Yamato SM310
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
Run “MENU”
“STERILIZE” Procedure
“STERILIZE” procedure
Check the following:
Is the drain valve closed? Pg.10
Is the drain filter in place? Pg.15
Is the bottom plate set? Pg.15
Is the bottle set? Pg.16
Is water in chamber? Pg.17
Is the lid closed? Pg.18
◇Turn on the main circuit breaker(located on the left side of
the unit).
◇Push the “MENU” key and select “STERILIZE”.
See P28
◇Set the sterilization temperature and time as you desire.
(See P25)
The main display blinks the set temperature.
The sub display blinks the set time.
The sterilize menu lamp comes on and off.
◇Push the “ENTER” key
Operation begins. The sterilization heater turns on.
The solenoid exhaust valve remains open. Air purges out of
the chamber. (Approximately 20-25 minutes)
The main display shows water temperature and heater
lamp.
The sub display goes off.
Sterilize lamp comes on and off.
Sterilize menu lamp blinks.
At the end of air purge, the solenoid exhaust valve closes,
and pressurization starts. (Approximately 20-25 minutes)
The main display shows the chamber temperature.
Sterilize lamp blinks.
Sterilize menu lamp is lit.

Related product manuals