19
Operational Precautions
Opening lid
When opening the lid, be sure that the
pressure gauge reads “0psi". Proceed to open
slowly. If the lid is opened at high pressure, hot
steam will spout out.
Drainage
After operation, the water in the chamber is
very hot. Please let the water cool before
draining. Place a drain pan less than 2 inches
in height under the drain valve or connect a
drain hose.
Never drain water during operation. If you
open the drain valve at high pressure, hot
water will spout out.
Water bottle window viewer
Check the water level window. If water level is
close to the upper limit indicated with the caution
label, be sure to drain some water before using.
OPEN
OPEN
Water Bottle
Window Viewer
The upper limit o
f
water level