X K 3 1 9 0-C 602
Chapter 4 Parameters Setting and Calibration
I. Parameters setting
The indicator has 5 groups of parameters setting condition, which are as follows:
【SET 0】: inquiry parameters;
【SET 1】:general parameters;
【SET 2】: control parameters;
【SET 3】: calibration;
【SET 4】: calibration parameters.
Press 【setting】 button to enter , press 【↑】or 【↓】to select the parameter group,
then press【input】 button to enter the setting of relevant parameter.
Some setting can go into effect after re-startup and power on.
Attention: when it is necessary to modify the calibrated parameters, the calibration
switch shall be turned on. Or else, the modification cannot be carried out and warning
“Error 7” will be prompted.
The contents of parameter table 【SET 0】 and【SET 2】 depend on the parameter
13 A in parameter table 【SET 1】.
Initialization of parameters
During the calibration process of (【SET 3】), input “123456” (ignore the decimal) while
inputting the loaded weight, press【input】 button, the indicator will carry out initialization of
parameters, all calibration parameters and working parameters will go back to the factory
setting. Under normal situation, special attention should be paid when using this function.
Explanation of button in parameters setting
【←】or【→】 change the currently flickering digit;
【↑】or 【↓】modify the value (parameter) of the currently flickering digit;
【esc】 do not save the current modified parameter, go back to the weighing status.
【input 】 if a parameter is modified, pressing this button will save the parameter and enter the
[ 0]