X K 3 1 9 0-C 602
8
[******]
[CALEnd
]
Display current load
weight
The calibration is done and
the calibration status is
quitted. Toggle the
calibration switch to OFF
and press【Input】to go back
to normal weighing status.
III. Checking of calibration data
Prompt: The calibration switch should be opened when the calibration data is modified,
otherwise it cannot be saved.
Press 【Setting】and 【↓】 to show . Press 【Input】, the indicator will enter
the data checking status, please see table 4-7.
Table 4-7 Calibration Data Checking
Parameter
Parameter
Display
Parameter Instruction
Operation Instruction
1
[ *]
[dC ]
Decimal Digits(0-3)
Press 【 Input 】 after
modifying the
parameter
2
[ **]
[e ]
Division(1/2/5/10/20/50)
Press 【 Input 】 after
modifying the
parameter
If the decimal digit is
not zero, 10, 20, 50 can
not be selected.
3
[******]
[F ]
Max. Weighing
Press 【 Input 】 after
modifying the
parameter
After modifying the
max. weighing
4
[******]
[0_Ad ]
Zero AD code
Press 【 Input 】 after
modifying the
parameter
5
[******]
[bL ]
CAL Coefficient
Press 【 Input 】 after
modifying the
parameter
6
[******]
[noLine]
Nonlinear Correction
Value(Note 1)
(-1~+1) Unit:%Max
Press 【 Input 】 after
modifying the
parameter
7
[ *]
[PZL ]
Tare
Cannot be modified
[ 4]