10-2
IM 04P02B01-01E
Code Message Explanation/Countermeasures
72 MATH channel is turned off
81 All space or ‘quit’ string cannot be specified.
86 The key-lock release password is incorrect. Enter the correct password.
87 This key is locked.
91 Password is incorrect. Enter the correct password.
100 IP address doesn’t belong to class A, B, or C.
101 The result of the masked IP address is all 0s or 1s.
102 SUBNET mask is incorrect. Set a correct subnet mask.
103 The net part of default gateway is not equal to Set the correct default gateway.
that of IP address.
151 This action is invalid during calculation.
160 This action is invalid during chart end.
161 This action is invalid during pen hold.
162 Cannot set an number for a skipped data.
163 This action is invalid during record.
164 This action is invalid during manual printing.
165 This action is invalid during list printing.
166 This action is invalid during setup list printing.
167 This action is invalid during chart feed.
169 This action is invalid during ribbon hold.
Operation Errors
Code Message Explanation/Countermeasures
232 There is no available data. There is no data for periodic printout or data for calculating TLOG
when the timer expired.
Communication Errors
Code Message Explanation/Countermeasures
300 Command is too long.
301 Too many number of commands delimited with ’;’. Keep the number of commands separated by sub delimiters
under 10.
302 This command has not been defined.
303 Data request command can not be enumerated with
sub-delimiter.
350 Command is not permitted to the current user level.
351 This command cannot be specified in the current
mode.
352 The option is not installed.
353 This command cannot be specified in the current
setting.
354 This command is not available during calculation.
390 Command error.
391 Delimiter error.
392 Parameter error.
393 No permission.
394 No such connection.
395 Use “quit” to close this connection. Attempted to disconnect its own connection.
396 Failed to disconnect.
397 No TCP control block. The control block of the specified connection cannot be found.
10.1 A List of Error Messages

Table of Contents

Other manuals for YOKOGAWA 436101

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA 436101 and is the answer not in the manual?

YOKOGAWA 436101 Specifications

General IconGeneral
BrandYOKOGAWA
Model436101
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals