Preparing for Measurement
3-13
IM AQ6375E-02EN
1
2
3
4
5
6
App
Index
You can also shut down the instrument using panel keys and soft keys.
1.
Press SYSTEM.
2.
Tap the More three times.
3.
Tap the Shut Down. The Shutdown screen is displayed.
4.
Tap the Yes. Shut down begins.
5.
After the POWER switch changes from green to orange, turn OFF the MAIN POWER switch
on the rear panel of the instrument.
Note
If for some reason the instrument fails to shut down normally, hold down the POWER switch for
approximately four seconds or longer to force standby mode. Note that the operating system configuration
file will not be backed up, possibly resulting in malfunctions upon start up the next time the instrument is
turned on.
Screen when the instrument was not shut down
If the shutdown procedure was not performed after the previous session, the following message
appears after start up. Failure to properly shut down the instrument can result in damage to the
monochromator. When turning OFF the power, always perform the shut down procedure.
Press any key to clear this message.
3.3 Turning the Power ON/OFF