Preparing for Measurement
3-19
IM AQ6375E-02EN
3
For Laser Type Light Sources
5.
Tap the External Laser. A screen for specifying the wavelength of the external light source is
displayed.
6.
Select the laser wavelength value using the rotary knob or the arrow keys. The allowed
wavelength range is 1200–2400 nm.
7.
Press ENTER. The wavelength value is set.
8.
Tap the Execute. Wavelength calibration is executed. The previous screen is displayed again
after the calibration process ends.
9.
Tap the Cancel during wavelength calibration to cancel the wavelength calibration process.
Enter the value.
Select from presets.
3.5 Wavelength Calibration