No. Message Cause of Warning
No. 200 - 299: Warnings for hardware failures
200 Fan motor stopped The fan motor (main unit) stopped. Automatically shuts down after
10 seconds from occurrence.
201 Calibration data failed Started up in emulation mode because there was a problem with
the instrument’s calibration data.
204 Calibration data failed There is an error in the instrument’s calibration data.
205 Internal communication error An abnormality occurred during instrument-internal communication.
206 Internal communication error An abnormality occurred during instrument-internal communication.
207 Abnormal internal temperature Internal temperature is abnormally high. Automatically shuts down
after 10 seconds from occurrence.
210 Internal Temparature warning Warning for rising internal temperature.
211 Auto offset error Obtained an abnormal value during AUTO OFFSET operation.
212 Auto offset error Obtained an abnormal value during AUTO OFFSET operation.
213 Auto temperature control error Problem with the light detector temperature control.
214 Measurement sequence error The sweep stopped because measurement sequence fell into
disorder during the sweep.
215 System optimization required System optimization is required. Restart the system.
220 Boot sequence error Started up in emulation mode because an abnormality occurred
upon start-up
221 Boot sequence error Started up in emulation mode because an abnormality occurred
upon start-up
222 Emulation mode. Entered emulation mode because an abnormality occurred.
223 Boot sequence error! Started up in emulation mode because an abnormality occurred
upon start-up
224 Internal communication error An abnormality occurred during instrument-internal communication.
225 Internal communication error An abnormality occurred during instrument-internal communication.
226 Internal communication error An abnormality occurred during instrument-internal communication.
227 Internal communication error An abnormality occurred during instrument-internal communication.
228 Memory allocation error Data memory initialization failed.
230 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
231 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
232 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
233 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
234 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
235 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
236 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
237 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.