System Errors (900 to 999)
Code Message Chapter or
Section
901 Failed to backup setup data.
Will initialize.
Backup battery may be low.
–
903 The USB device's power consumption exceeded the capacity of the USB hub. –
904 Lower the sample rate or reduce the number of measuring channels. Chapter 2
1
905 Lower the sample rate or reduce the number of measuring channels. Chapter 2
1
906 Fan stopped.
Maintenance service is required.
It will shutdown.
–
907 Internal temperature is too high.
Maintenance service is required.
It will shutdown automatically.
–
909 SD Card is full. 6.10
1
910 Key protect is enabled.
To release the protection, press the PROTECT key or enter the password.
20.5
1
912 Fan for CPU stopped.
Maintenance service is required.
It will shutdown automatically.
–
913 LCD BackLight Failure.
Maintenance service is required.
–
915 It installed the module which cannot support by this machine. –
917 Hardware configuration error occurred.
Restart this machine.
If it occurred again, maintenance service is required.
2.9
918 Error occurred while ImageFile process. –
919 Key operate not available while system error occurred –
920 Firmware overwriting error occurred. –
922 GPS module does not work. 2.22
923 Setting was changed.
Will initialize.
–
930 Slot 1 EEPROM error. –
931 Slot 2 EEPROM error. –
938 Internal flash write error. –
939 Internal flash verify error. –
940 Because abnormal temperature was detected, then it will shut down. –
941 Because an unspected battery was detected, then it will shut down. 2.5
942 Because USB overcurrent was detected, the USB power was turned off.
Please disconnect the connected USB device and restart it.
2.9
943 Because GPS overcurrent was detected, the GPS power was turned off.
Please disconnect the GPS unit and restart it.
2.22
944 Because module insertion/removal was detected, acquisition was stopped.
Please restart it.
2.9
945 Low Battery.
Please connect to another power supply immediately.
2.6, 2.7
946 Because a battery residual quantity disappeared, It will shutdown. 2.9
947 Battery voltage is low.
Please connect to another power supply immediately.
2.6, 2.7
1 User’s Manual, IM DL350-02EN
Note
If servicing is required, first see if initializing the instrument fixes the problem.