Code Message Chapter or
Section
850 Cannot be set when the trace is set to a frequency module. –
852 Cannot set Math to OFF while FFT Window ON. 10.1
1
853 Cannot select this trace because it already selected. –
854 Because a record length is too long, it can't be set up by the present number of indication channels. Chapter 2,
1
3.1
1
855 Cannot change to such Record length while running.
Set the trigger mode to Auto and decrease T/Div to less than 100 msec/div to enable roll mode, or set
the trigger mode to Single or On Start.
–
856 Cannot Display setting to On.
This CH didn't acquisition to memory.
Chapter 2
2
858 Cannot set while action mode is on. 3.1
1
868 Cannot be specified because characters in the JIS level-2 kanji set are included.
Create the file on the local drive, and then copy it to the network drive.
–
869 Cannot set while Go-Nogo mode.
Turn Off Go-Nogo mode first.
Chapter 13
1
870 All sub-channel inputs are off.
At least one inputs should on.
2.3,
1
2.5
1
871 No effective channel for Math Setup. –
875 Cannot change or START when accumulate is ON.
Turn OFF accumulate first.
5.3
1
879 Cannot set GoNogo mode while Math or FFT Window is On. Chapter 9,
1
chapter 10
1
881 Cannot set for CH which ValueType is Float while running. –
883 Cannot set Input to ON with limit of memory capacity. 2.10,
1
2.11,
1
2.12
1
885 Zoom rate is limited while installed a module which has sub-channels. –
890 LIN Monitor does not support Float data. 2.11
1
892 Cannot set Math operation while FFT window is open. 10.1
1
893 Cannot set that item.
It cannot be measured for Logic module.
8.1
1
894 Cannot set DC Offset Cancel to ON due to following case.
- Coupling is not DC.
- Not execute DC Offset Cancel.
2.2,
1
3.6
895 The password entered the first time is different from the password entered the second time.
Reenter the password for the second time.
–
896 It is not possible to set the subchannel sample interval to be 1 second or more.
(sample rate to be 1S/s or less.)
3.1,
1
3.2
1
897 This function cannot be used when
the recorder mode is set.
–
898 Cannot change when Internal Clock is active. –
1 User’s Manual, IM DL350-02EN
2 Features Guide, IM DL350-01EN
5.2 Messages and Corrective Actions