Setup Errors (800 to 899)
Code Message Chapter or
Section
800 Illegal date-time.
Set the correct date and time.
–
801 Cannot set these file name.
- Over 32 characters.
- Contains character which are not allowed.
- Inhibit MS-DOS file name.
Enter an other file name.
6.3
1
802 Cannot set while recording. 3.1,
1
3.2
1
804 Cannot change settings during GO/NO-GO.
Stop the GO/NO-GO (Stop the Acquire) .
Chapter 13
1
806 Cannot be changed when trigger A is not X.
Set the state of the channel corresponding to
condition A to 'X'.
4.7
1
to 4.12
1
808 Cannot change when Channel Display is OFF or Math settings are invalid.
Set the channel display ON or make appropriate Math settings.
Chapter 2,
1
Chapter 9
1
809 Cannot change when External Clock is active. 3.1,
1
3.2
1
814 Cannot change settings while SD Card recording.
Stop SD recording.
–
815 Cannot change settings during Action mode.
Stop the Action.
3.1,
1
3.2
1
816 Cannot set the channels which do not have modules installed. Chapter 2,
1
5.6
2
817 Cannot Set or Execute. –
818 If the trigger mode is set to Single or OnStart, the acquisition mode cannot be set to Average. 3.1
1
819 If the acquisition mode is Average, the trigger mode cannot be set to Single or OnStart. 4.1
1
820 The acquisition mode cannot be set in the current record length. –
822 Cannot be configured or executed during the search operation. Chapter 16
1
824 The record cannot be selected. 17.1
1
825 History record does not exist. 17.1
1
826 Cannot be configured or executed while computation is in progress.
Aborted when history display mode is set to One.
17.1
1
827 Cannot be configured or executed while updating the history all display.
Aborted when history display mode is set to One.
17.1
1
829 Zones cannot be edited in the following cases:
* When the main window is not displayed.
* When the relevant waveform is not displayed.
13.1
1
830 The zone waveform does not exist. 13.1
1
831 The zone is being edited.
To perform other operations, exit zone editing.
13.1
1
832 Zones determination is not possible in the following cases:
* When the main window is not displayed.
* When the relevant waveform is not displayed.
* When the zone waveform does not exist.
Chapter 2,
1
10.1,
1
11.2,
1
14.2,
1
15.2
1
833 Processing statistics.
To perform other operations, abort the statistical processing.
8.2
1
834 There are no channels to be analyzed,
or you specified a channel that can not be set.
–
835 Cannot be set when the acquisition mode is set to average. 3.1
1
836 Cannot be changed when VScale is SPAN. 2.2
1
837 Cannot be set during SD Card recording. –
840 Cannot be set to a range of 20 sec/div to 3 day/div during roll display. 3.1,
1
3.2
1
841 Cannot be set because there are too many display channels at the current record length.
Shorten the record length.
3.1
1
845 Cannot change the History parameter when accumulate is ON.
Turn OFF accumulate first.
5.3
1
847 Cannot set On this module. –
848 Settings can not be entered for channels on which no strain module is mounted. –
849 Setting error. –
5.2 Messages and Corrective Actions
1 User’s Manual, IM DL350-02EN