2.10  Connecting Probes
•  720250 (HS10M12)
  Maximum input voltage (at a frequency of 1 kHz or less)
•  In combination with an safety probe (10:1) 700929 or (100:1) 701947 or a wide 
temperature range type (10:1) 702902
1
  800 V (DC + ACpeak)
•   Safety cable (1:1) (combined with 701901+701954)
3
  200 V (DC + ACpeak) (as a value that meets the safety standard)
  250V (DC + ACpeak) (maximum allowable voltage, as a value that does not damage the 
instrument when applied)
•  Direct input (cable that does not comply with the safety standards)
5
  42 V (DC + ACpeak)
  Maximum rated voltage to ground (at a frequency of 1 kHz or less)
•  In combination with a safety probe (10:1) 700929, (100:1) 701947, or wide temperature 
range type (10:1) 702902,
2
 or a safety cable (1:1) (701901+701954)
4
  400 Vrms (CAT II)
  Depending on the frequency of the signal applied to the L input, the maximum rated 
voltage to ground of the L input of the safety probe and safety cable is as shown in the 
following graph.
   
10
100
1 10 100 1000 10000 100000
1
•  Direct input (cable that does not comply with the safety standards)
6
  42 V (DC + ACpeak) (CAT II, 30 Vrms)