Page #255 background image
22-2
IM DLM3054-01EN
Default Settings (DEFAULT SETUP)
You can reset the instrument settings to their factory default values. This feature is useful when you want to
cancel all the settings that you have entered or when you want to redo measurement from scratch. Default setup
refers to the act of resetting the instrument settings to their factory default values.
Items That Cannot Be Reset
The following settings cannot be reset.
Date and time settings
Communication settings
Language setting (English or Japanese)
Measured value font size setting
Touch panel feature on/off
Undoing Default Setup (Undo)
If you perform default setup by mistake, you can undo it by pressing the Undo soft key. However, you cannot
undo the reset operation if you clear the Undo menu item by switching to a different setup menu or pressing the
ESC key.
To Reset All Settings to Their Default Values
While holding down the RESET key, turn the power switch on. All settings except the date and time settings (display
on/off setting will be reset) and the setup data stored in internal memory will be reset to their factory default
values.
Snapshot (SNAP SHOT)
Retains the currently displayed waveforms on the screen. This feature allows you to update the display without
having to stop waveform acquisition. It is a useful feature when you want to compare waveforms.
Snapshot waveforms are displayed in white and moves behind the normal waveforms.
You cannot perform the following operations on snapshot waveforms.
Cursor measurement, automated measurement of waveform parameters, zoom, or computation
You can save load snapshot waveforms.
22 Other Features

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM3054 and is the answer not in the manual?

YOKOGAWA DLM3054 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM3054
CategoryTest Equipment
LanguageEnglish

Related product manuals