8-2
 
Refer to Appendix D to learn more about report format options for the 2700 SELECT. For
troubleshooting, or even daily log information, the "detail" report format is preferable.
The Detail Report provides a complete description of the sensors for a calibration or sample.
Information for both the black and white probes, as well as the temperature probe, is included.
You may access the report format level without exiting RUN or STANDBY modes. Press
[MENU] on the keypad to display the Select setup menu level. Press [4] for Report to change or
confirm sample and calibration formats.
Listed below are example printouts and explanations of the Detail format information.
Sample Report (Detail) Calibration Report (Detail)
-----Sample Report----- ==CALIBRATION REPORT==
ID: 123456789-05-03 B:Lactate * Unstable *
B:L-Lactate 4.82 g/L IB current 1.79 nA
IB current 2.11 nA PL current 11.63 nA
PL current 118.65 nA FB current 1.51 nA
Slope 4.69 nA/min *Base shift -2.65 %
End Point 30 Sec Slope 1.42 nA/min
W:Dextrose 8.3 g/L End Point 30 Sec
IB current 0.88 nA *Cal shift 3.64 %
PL current 58.96 nA W:Dextrose 10.0 g/L
Slope 2.39 nA/min IB current 0.88 nA
End Point 30 Sec PL current 5.81 nA
Temperature 26.28 °C FB current 0.76 nA
Sample size 25 uL Base shift -1.95 %
Fri 02/22/98 08:12:34 Slope 0.09 nA/min
YSI 2700D - 98 01234 End Point 30 Sec
----------------------- Cal shift -0.10 %
Temperature 26.33 °C
Sample size 25 uL
Thu 02/22/98 14:55:10
YSI 2700D - 98 01234
=======================