EasyManua.ls Logo

YSI 2700 SELECT - L-Glutamine

YSI 2700 SELECT
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-15
 #
This is a direct reading of glutamine in solution at the enzyme sensor. Two enzymes are co-
immobilized in the YSI Glutamine Membrane: Glutaminase and Glutamate Oxidase.
L-Glutamine
Glutaminase
L-Glutamate + NH
3
L-Glutamate + O
2
Glut Oxidase
H
2
O
2
+ α-Ketoglutarate + NH
3
Through this chain of reactions the amount of hydrogen peroxide liberated is directly
proportional to the amount of glutamine.
System Buffer YSI 2357
Calibrator Std YSI 2736 (5.00 mmol/L)
Linearity Std. YSI 2737 (8.00 mmol/L)
Black Probe YSI 2735 Glutamine Membrane
White Probe n/a *
Sample Size 20 µL
Cal Station Station #2
Black Probe Parameters
Chemistry Glutamine
Unit of Conc mmol/L
Cal value 5.00
End Point 30 sec
* If you have a dual channel unit, either Black or White, or both Black and White probes can be
configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water or 2357 buffer.
The sample must be
glutamate-free
, or at least contain levels low enough not to interfere with
the glutamine reading. Since the glutamine membrane contains glutamate oxidase, glutamate
will produce a probe signal. See Section 4.17 to measure glutamate and glutamine
simultaneously.

Table of Contents

Related product manuals