D-1
When you setup your YSI 2700 SELECT from the Setup Menu, you have the option to print the
calibration and sample results in "brief" or "detail" format. You also have the option not to print
calibration values. Review Section 5.3, Setup, specifically choice 4-Report in the Setup
selection menu. For more specific information on the “detail” printout, refer to Section 8,
Troubleshooting.
Example printouts of both formats are presented below. These examples show printed results of
samples and calibrations. In the first Sample Report, the ID# “123456789-05-03” indicates that
the third of five replicates is being reported for the sample identified by the nine digit number
“123456789”.
Sample Report (Detail) Calibration Report (Detail)
-----Sample Report----- ==CALIBRATION REPORT==
ID: 123456789-05-03 B:L-Lactate* Unstable *
B:L-Lactate 4.82 g/L IB current 1.79 nA
IB 2.11 nA PL current 11.63 nA
PL current 18.65 nA FB current 1.51 nA
Slope 0.69 nA/min *Base shift -2.65 %
End Point 30 Sec Slope 1.42 nA/min
W:Dextrose 5.38 g/L End Point 30 Sec
IB current 0.88 nA *Cal shift 3.64 %
PL current 15.96 nA W: Dextrose 2.50 g/L
Slope 0.39 nA/min IB current 0.88 nA
End Point 30 Sec PL current 5.81 nA
Temperature 26.28 °C FB current 0.76 nA
Sample size 25 uL Base shift -1.95 %
Fri 02/27/98 08:12:34 Slope 0.09 nA/min
YSI 2700D - 98 01234 End Point 30 Sec
----------------------- Cal shift -0.10 %
Temperature 26.33 °C
Sample Report (Brief)
Sample size 25 uL
Thu 02/26/98 14:55:10
ID# 123456789 YSI 2700D - 98 01234
Replicate 1 of 3 =======================
-----Sample Report-----
ID: 123456789-03-01
B:L-Lactate 0.82 g/L
Calibration Report (Brief)
W: Dextrose 5.31 g/L
Wed 02/25/98 13:22:56 ==CALIBRATION REPORT==
----------------------- B:L-Lactate 16.68 nA
W: Dextrose 12.81 nA
Mon 02/23/98 17:12:09
=======================