EasyManua.ls Logo

YSI 2700 SELECT - Sucrose

YSI 2700 SELECT
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-7
$ 
This is a direct reading of sucrose in solution at the enzyme sensor. Three enzymes are co-
immobilized in the YSI Sucrose Membrane: Invertase, Mutarotase, and Glucose Oxidase.
Sucrose + H
2
O
Invertase
α-D-Glucose + (Fructose)
α-D-Glucose
Mutarotase
β-D-Glucose
β-D-Glucose + O
2
Glu Oxidase
H
2
O
2
+ D-Glucono-δ-Lactone
Through this chain of reactions the moles of hydrogen peroxide liberated is directly
proportional to the moles of sucrose.
System Buffer YSI 2357
Calibrator Std YSI 2780 (5.00 g/L Sucrose)
Linearity Std YSI 2778 (25.0 g/L Sucrose)
Black Probe YSI 2703 Sucrose Membrane
White Probe n/a *
Sample Size 25 µL
Cal Station Station #1 Cal Well
Black Probe Parameters
Chemistry Sucrose
Unit of Conc g/L
Cal value 5.00
End Point 30 sec
* If you have a dual channel unit, either Black or White, or both Black and White probes can be
configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water or 2357 buffer.
»
The sample must be
dextrose-free
, or at least contain levels low enough not to
interfere with the sucrose reading. Since the sucrose membrane contains glucose
oxidase, dextrose will produce a probe signal. See Section 4.16 to measure dextrose
and sucrose simultaneously.

Table of Contents

Related product manuals