EasyManua.ls Logo

YSI 2700 SELECT - Choline

YSI 2700 SELECT
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-11
) 
This is a direct reading of choline in solution at the enzyme sensor. The enzyme Choline
Oxidase is immobilized in the YSI Choline Membrane.
Choline + 2O
2
Chol Oxidase
2H
2
O
2
+ Betaine
System Buffer YSI 2357
Calibrator Std YSI 2772
Linearity Std YSI 2773
Black Probe YSI 2771(Choline Oxidase Membrane)
White Probe n/a *
Sample Size 25 µL
Cal Station Station #1 Cal Well
Black Probe Parameters
Chemistry Choline
Unit of Conc mg/L
Cal value 175
End Point 30 sec
* If you have a dual channel unit, either Black or White, or both Black and White probes can be
configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water or 2357 Buffer.
»
Although the YSI 2772 Choline calibrator solution is prepared using choline bitartrate,
the concentration is expressed as mg/L of choline cation. If you prefer to express the
sample as a salt of choline, you must enter an “adjusted” calibration value when
configuring your instrument measurement parameters.
For example, to express results as a choline hydroxide value, program the calibrator
value as 204 mg/L when using the YSI 2772 Choline Standard calibrator solution. Your
results will now reflect the concentration of choline hydroxide in the sample.
Assumptions and calculations are as follows:
choline hydroxide FW = 121.2
free choline FW = 104.2
(121.1 ÷ 104.2) x 175 mg/L = 203.55 = 204 mg/L (rounded up)
Equivalent values for other choline salts may be calculated in a similar manner.
Potential Interferences. Tests with equimolar concentrations of various vitamins, and other
ingredients of nutritionals, showed little or no interference. On a scale where Choline = 100%,
those substances that produced more than 0.5% response were Riboflavin (1.1%), Pyridoxine
(1.5%), Ascorbic Acid (0.8%), and Thiamine (1.3%). Vanillin, an artificial flavor ingredient,
may produce interference levels exceeding 10% (mole/mole), and should be tested separately
for your individual formulations if known to be present.

Table of Contents

Related product manuals