YSI 2700 SELECT - 8.2 Troubleshooting Chart

YSI 2700 SELECT
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-4
 
SYMPTOM:
ERROR: Baseline
POSSIBLE CAUSE: Pinched, leaking or disconnected tube.
ACTION: Fix or replace tubing.
SECTION: 7.6
POSSIBLE CAUSE: Sipper misaligned.
ACTION: Check Sipper alignment.
SECTION: 2.6
POSSIBLE CAUSE: Buffer pump not operating properly.
ACTION: Replace tubing.
SECTION: 7.6
POSSIBLE CAUSE: Stir bar not in chamber.
ACTION: Disassemble chamber and reinstall stir bar.
SECTION: 7.3
POSSIBLE CAUSE: Newly installed enzyme membrane.
ACTION: Enter probe diagnostics and check probe currents.
SECTION: 2.6
POSSIBLE CAUSE: Newly installed probe.
ACTION: Enter probe diagnostics and check probe currents.
SECTION: 2.6
POSSIBLE CAUSE: Power disruption.
ACTION: Enter probe diagnostics and check probe currents.
SECTION: 2.6
POSSIBLE CAUSE: Failing enzyme membrane.
ACTION: Perform daily operational checks and replace membrane(s)
if necessary.
SECTION: 3.4
POSSIBLE CAUSE: Enzyme Probe surface(s) fouled.
ACTION: Clean probe surface(s).
SECTION: 7.5
POSSIBLE CAUSE: Temperature Probe (auxiliary electrode) fouled.
ACTION: Clean probe surface.
SECTION: 7.5
POSSIBLE CAUSE: Sample may contain an interfering substance.
ACTION: Attempt to confirm interference.
SECTION: 4.12
SYMPTOM:
ERROR: Empty Buffer Bottle
POSSIBLE CAUSE: Low buffer level.
ACTION: Refill the buffer.
SECTION: 2.3