8-5
SYMPTOM:
ERROR: Empty Buffer Bottle
(Continued)
POSSIBLE CAUSE: Level sensor cable not plugged in.
ACTION: Plug in cable.
SECTION: 2.3
SYMPTOM:
ERROR: Empty Calibrator Bottle
POSSIBLE CAUSE: Low calibrator level.
ACTION: Install new calibrator.
SECTION: 2.3
POSSIBLE CAUSE: Level sensor cable not plugged in.
ACTION: Plug in cable.
SECTION: 2.3
SYMPTOM:
ERROR: Full Waste Bottle
POSSIBLE CAUSE: High waste level.
ACTION: Empty bottle.
SECTION: ---
POSSIBLE CAUSE: Level sensor cable in contact with instrument housing or
another conductive body.
ACTION: Reroute cables.
SECTION: 2.3
SYMPTOM:
ERROR: Motor Failure
POSSIBLE CAUSE: One of the motors is jammed.
ACTION: Enter motor diagnostics and cycle the suspected motor.
SECTION: 5.4
POSSIBLE CAUSE: Dirty drive screw on sipper mechanism.
ACTION: Clean and lightly oil.
SECTION: 7.11
POSSIBLE CAUSE: Worn sipper pump seals.
ACTION: Replace seals.
SECTION: 7.10
SYMPTOM:
ERROR: Overrange
POSSIBLE CAUSE: Sample concentration too high, resulting in high probe nA.
ACTION: Dilute sample or adjust sample size down and repeat.
SECTION: 4.3
SYMPTOM:
ERROR: Printer Failure Detected
POSSIBLE CAUSE: Printer paper or roll jam.
ACTION: Open paper well and clear obstruction. Reset instrument,
if required.
SECTION: 2.6