2-18
Choosing Yes maintains level sensing in the Buffer, Calibrator, and Waste bottles. Press
[ENTER] to return to General setup.
General setup: 1-Date/Time 2-Contrast
3-RS232 4-Radix 5-Serial# 6-LevelSensor
You have now completed General setup. Press [0] to return to Select setup menu level.
Select setup: 1-General 2-MeasParameter
3-RunMode 4-Report 5-PrntSetup 6-Default
Press [2] for MeasParameter.
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White
Press [1] for SampleSize.
Enter sample size in microliter
> 25
The default setting is 25 microliter (µL). The options range from 5 to 65 µL. If you enter a
number outside of this range, the display will show you the limits. Press [ENTER] to confirm
25 µL and return to the Measurement parameter setup menu level.
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White
Press [2] for CalMethod.
Select calibration method: One Station
1-One Station 2-Two Stations
Press [ENTER] to confirm One Station, since both dextrose and L-lactate are in one calibrator
bottle and reside in one location.
Enter Cal Station #
> 1
Press [ENTER] to confirm that the calibrator solution (YSI 2776) is installed at Station #1 (Cal
Well of the Sample Chamber). The display now returns to the Measurement parameter setup
menu level.
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White