3-6
The sample will be aspirated and the Sipper will return to the Sample Chamber. Do not move
the sample container until the Sipper has returned the Sample Chamber.
Aspirating sample...
The results are then displayed and printed as shown above.
Again compare the value reported by the instrument with the value on the label of the YSI 1531
Standard (9.00 g/L dextrose). Repeat calibration and sampling if the result is not within
specified tolerance limits of ±5% as described above. If your reading is still outside this range,
refer to Linearity Test below.
Figure 3.4
The Manual Station