EasyManua.ls Logo

YSI 2700 SELECT - Page 53

YSI 2700 SELECT
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-4
Calibration Method (2-CalMethod).
In most chemistry setups, you will select “One Station”,
and the calibration station # will be “Station #1” (Calibration Well). However, you have options
to configure this in other ways.
»
Select One Station, but assign Cal Station #2. Station #2 is the Test Tube Holder
Station. For example, you may place a 500 mg/L dextrose solution in a test tube and
calibrate in this manner. If you choose this option, the sampling position is normally
Station #3 (Manual Station). With accessories, Station #4 (Turntable), or Station #5
(Monitor) may be used.
»
If you are running a dual chemistry where a potential interference is involved, you may
select “Two Station”. You will then be prompted to assign calibration stations for both
black and white probes. For example, you could install a sucrose membrane on one
probe and a dextrose membrane on the second probe. You could place sucrose
calibrator in the test tube position (#2) and dextrose at the Cal Well (#1). In this setup,
the instrument is programmed to measure the “dextrose reading” at the sucrose
membrane and automatically subtract it.
Black Probe Parameter (3-Black).
In this menu you choose parameters that correspond to the
chemistry of interest at the Black Probe. Be certain that you install the correct enzyme
membrane type. The selections include:
»
Chemistry (10 selections, plus “none”)
»
Unit of Concentration (g/L, mg/L, mmol/L, %(w/v))
»
Calibration value (must correlate to units chosen)
»
End Point (basically, the reaction time in the sample chamber)
For most standard applications, recommended parameter settings are outlined below under each
chemistry setup.
White Probe Parameter (4-White).
In this menu, you choose parameters that correspond to
the chemistry of interest at the White Probe. Be certain that you install the correct enzyme
membrane type. The selections include:
»
Chemistry (10 selections, plus “none”)
»
Unit of Concentration (g/L, mg/L, mmol/L, %(w/v))
»
Calibration value (must correlate to units chosen)
»
End Point (basically, the reaction time in the sample chamber)
For most standard applications, recommended parameter settings are outlined below under each
chemistry setup.

Table of Contents

Related product manuals