5-25
If the Sipper Pump plunger is at "home", i.e. extended, the SipperPump message should be ON;
if the plunger is in any position other than home, the message should be OFF.
If there is a test tube in the Test Tube Holder and it is in a vertical orientation, the "TestTube"
message should be ON; if not, then the message should be OFF.
Press [0] to return to Sensor diagnostic menu.
Sensor diagnostic
1-Temperature 2-LevelSensor 3-OptoSensor
Press [0] to return to the five diagnostic selections.
Select diagnostic
1-Motor 2-Pump 3-Probe 4-I/O 5-Sensor
Press [0] to return to Select instrument function.
Select instrument function
1-Service 2-Setup 3-Diagnostic
Press [0] to return to Main Menu display.
Please select instrument mode
[RUN] [STANDBY] [MENU]
NOTE
: You may press [MENU] from most levels of diagnostics to return directly to Main
Menu.
You now have had a complete look at the Diagnostic Menu. You may press [RUN] or
[STANDBY] to enter an operating mode, or remain in Main Menu indefinitely, however, note
that no automatic calibrations or sample chamber flushes will occur in this mode.