Operation
2.6
Sensor Menu
Push the key to access the sensor menu (Figure 22). Highlight a sub-
menu then push the
ENTER
key to view sub-menu options.
Pre-defined or user-selected sensor settings are noted within brackets ([]).
Figure 22 Probe (Sensor) menu
Use the Probe (
) key to access the Sensor menu and change sensor settings (if applicable), enable the measure-
ment units displayed on the Run screen, set Auto Stable parameters, change the sensor averaging mode, and if
equipped, turn on/off GPS.
Sensor Setup
Setup
The Sensor Setup menu will show all sensors connected to the instrument
(Figure 23). If a sensor is connected but is not listed on the Sensor Setup
menu (<None> displayed), check the sensor and cable connections.
Figure 23 Sensor Setup
Setup ODO
Setup ODO
Local DO: Enable or disable localized DO% measurements. When
enabled, the calibration value is set to 100% regardless of altitude or
barometric pressure. When enabled, an L will be shown next to DO% on
the run screen. DO mg/L measurements are unaffected when Local DO is
enabled (Figure 24).
LDS: Last Digit Supression (LDS) rounds the DO value to the nearest
tenth, e.g. 8.27 mg/L becomes 8.3 mg/L.
Sensor Cap Coefficients: The sensor cap coefficients must be updated
after sensor cap replacement. Update the sensor cap coefficients using
the coefficient sheet provided with the new sensor cap. Once updated,
the coefficients are saved to the ODO sensor and do not need to be
re-entered.
NOTE: The coefficients stay with the sensor even when used with
different handheld meters.
Figure 24 Setup ODO