Appendix A - DO% Calibration Values
Calibration Value Pressure
D.O. % in Hg mmHg kPa mbar
101% 30.22 767.6 102.34 1023.38
100% 29.92 760.0 101.33 1013.25
99% 29.62 752.4 100.31 1003.12
98% 29.32 744.8 99.30 992.99
97% 29.02 737.2 98.29 982.85
96% 28.72 729.6 97.27 972.72
95% 28.43 722.0 96.26 962.59
94% 28.13 714.4 95.25 952.46
93% 27.83 706.8 94.23 942.32
92% 27.53 699.2 93.22 932.19
91% 27.23 691.6 92.21 922.06
90% 26.93 684.0 91.19 911.93
89% 26.63 676.4 90.18 901.79
88% 26.33 668.8 89.17 891.66
87% 26.03 661.2 88.15 881.53
86% 25.73 653.6 87.14 871.40
85% 25.43 646.0 86.13 861.26
84% 25.13 638.4 85.11 851.13
83% 24.83 630.8 84.10 841.00
82% 24.54 623.2 83.09 830.87
81% 24.24 615.6 82.07 820.73
80% 23.94 608.0 81.06 810.60
79% 23.64 600.4 80.05 800.47
78% 23.34 592.8 79.03 790.34
77% 23.04 585.2 78.02 780.20
76% 22.74 577.6 77.01 770.07
75% 22.44 570.0 75.99 759.94
74% 22.14 562.4 74.98 749.81
73% 21.84 554.8 73.97 739.67
72% 21.54 547.2 72.95 729.54