Operation
View Calibration Record
→ View Calibration Record
Select View Calibration Record to show the stored sensor calibrations
(Figure 40).
Use the arrow keys to scroll through the calibration file data.
Delete Data
→ Delete Data
Enter the desired filter criteria, then select Delete Selected Data to
permanently delete the data (Figure 41).
Select Delete All Data to permanently delete all logged data from the
handheld.
Calibration Information
Information in each calibration record:
• Sensor calibrated
• Date/time stamp
• Sensor ID
• Sensor serial #
• Sensor software version
• User ID (optional)
• Probe ID (optional)
• User Fields #1 and #2 (optional)
• Calibration status
• Calibration value
• Temperature
Depending on the parameter, a calibration record may include additional
information such as the Conductivity cell constant, ODO gain, ORP offset,
and pH slope.
Figure 40 View GLP
Figure 41 Delete Data Filter