Calibration
Calibration Screen Layout
The calibration screen has the same basic layout for each parameter
(Figure 46).
Calibration value: This is the value the sensor will be calibrated to. The
Yellow Line on the graph corresponds to this value.
Accept Calibration: Select this to calibrate the sensor to the calibration
value.
Finish Calibration: This option is only available with multi-point
calibrations (i.e. pH, ISE, turbidity, PC, PE, and chlorophyll). Finishes the
calibration by applying previously accepted points.
Press ESC to Abort: Press the ESC key to leave the calibration. The sensor
will not be calibrated to any points. The last successful calibration will be
used.
Last Calibrated: View the date and time of the last successful sensor
calibration.
Actual Readings: This shows the current measurement value on the Run
screen. The White Line on the graph corresponds to this value. Observe
the White Line to ensure the measurement is stable before choosing
Accept Calibration.
Post Cal Value: This is the same as the calibration value. This will be the
measurement value in the current solution after the calibration is finished.
Figure 46 Layout of calibration screen
Calibration Setup (continued)