6 9650-0450-01 Rev. U
M Series Service Manual
WAR NI NG ! This unit can generate up to 4500 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the defibrillator.
Do not discharge the unit’s internal energy more than three times in one minute or damage to the unit may result.
Do not discharge a battery pack except in a Base PowerCharger
4x4
or compatible ZOLL Battery Charging/Testing unit.
Do not use the M Series in the presence of flammable agents (such as gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the unit
near the site of a gasoline spill may cause an explosion.
Do not use the unit near or within puddles of water.
Note: The M Series is protected against interference from radio frequency emissions typical of two-way radios and cellular phones (digital and
analog) used in emergency service/public safety activities. Users of the M
Series should assess the unit’s performance in their typical
environment of use for the possibility of radio frequency interference from high-power sources. Radio Frequency Interference (RFI) may be
observed as shifts in monitor baseline, trace compression, or transient spikes on the display.
Additional Reference Material
In addition to this guide, there are several other components to the ZOLL M Series documentation. They include:
Operators Guide - A comprehensive reference work that describes all the user tasks needed to operate the M Series.
Configuration Guide - Describes the M Series features and functions whose operation can be customized by authorized users.
Conventions
WARNING! Warning statements describe conditions or actions that can result in personal injury or death.
Caution Warning statements describe conditions or actions that can result in personal injury or death.
Note: Notes contain additional information on using the defibrillator.

Table of Contents

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules