E
If the unit does not function properly, remove the
battery and reinsert it as described above.
Battery Charging
E
The unit contains a powerful lithium-ion rechargeable
battery. The battery does not have any memory
effect and can therefore be recharged at any
time by placing it in the charger (see the section
describing the battery power level display on the
handpiece).
E
The battery can also be charged in the charger
independently of the handpiece.
E
Place the handpiece on the charger for approx.
1.5 hours before the first use, in order to charge the
new battery completely for the first time.
-
The status LED showing the operating status of
the charger flashes green during the charging
operation; see below “Operating Status Display
of the Charger.
Operating Status Display of the Charger
Power Level Display of the Handpiece
Sealing
Status LED Operating status
Handpiece/battery Handpiece/battery
outside the charger
in the charger
Steady green light Charger ready for Charging has
operation been completed
Flashes green –– Battery is being
charged
Steady red light Charge contact Charge contact
pins are wet pins are wet
Flashes alternately Malfunction in the Problem during
red and green charger charging
Power level LED Operating status
Handpiece outside Handpiece in the
the charger charger
Steady green light Battery is charged, Not possible,
handpiece ready handpiece is in
for operation sleep mode
Steady red light Warning for low Not possible,
battery capacity, handpiece is in
remaining capacity sleep mode
adequate typically
for 5 10-sec cycles
Flashes red Battery fully Problem in
discharged, charging, battery
exposure cycle is defective or
will be completed cannot be charged
or, if in continuous
mode, stopped
Operation