A56 User Guide
Register Modbus
number
Parameter Function Datatype DataRange
90 40090 RESERVED–C₂H₄ RoC (ppm/
month)
return 0 signed 16-bit binary number
91 40091 C₂H₆ status 16-bit binary map see <Gas
status> table
92 40092 C₂H₆ (ppm) signed 16-bit binary number
93 40093 C₂H₆ RoC (ppm/day) signed 16-bit binary number
94 40094 RESERVED–C₂H₆ RoC (ppm/
week)
return 0 signed 16-bit binary number
95 40095 RESERVED–C₂H₆ RoC (ppm/
month)
return 0 signed 16-bit binary number
96 40096 TDCG status 16-bit binary map see <Gas
status> table
97 40097 TDCG (ppm) signed 16-bit binary number
98 40098 TDCG RoC (ppm/day) signed 16-bit binary number
99 40099 RESERVED–TDCG RoC (ppm/
week)
return 0 signed 16-bit binary number
100 40100 RESERVED–TDCG RoC
(ppm/month)
return 0 signed 16-bit binary number
101 40101 C₃H₆ status 16-bit binary map see <Gas
status> table
102 40102 C₃H₆ (ppm) signed 16-bit binary number
103 40103 C₃H₆ RoC (ppm/day) signed 16-bit binary number
104 40104 RESERVED–C₃H₆ RoC (ppm/
week)
return 0 signed 16-bit binary number
105 40105 RESERVED–C₃H₆ RoC (ppm/
month)
return 0 signed 16-bit binary number
106 40106 C₃H₈ status 16-bit binary map see <Gas
status> table
107 40107 C₃H₈ (ppm) signed 16-bit binary number
108 40108 C₃H₈ RoC (ppm/day) signed 16-bit binary number
109 40109 RESERVED–C₃H₈ RoC (ppm/
week)
return 0 signed 16-bit binary number
110 40110 RESERVED–C₃H₈ RoC (ppm/
month)
return 0 signed 16-bit binary number
111 40111 Oil temperature status 16-bit binary map see <Gas
status> table
112 40112 Oil temperature (moisture
sensor) (Celsius)
signed 16-bit binary number
113 40113 Oil temperature RoC (ppm/day) signed 16-bit binary number
114 40114 RESERVED–Oil temp. RoC
(ppm/week)
return 0 signed 16-bit binary number
115 40115 RESERVED–Oil temp. RoC
(ppm/month)
return 0 signed 16-bit binary number