• Output on the Warning/Alarm channels are given by integer values between 4 - 20 as defined in the
user manual:
o 4 - 10 indicates various alarms conditions
o 12 - 18 indicates various warning conditions
o 20 indicates status is good
• Output on the gas concentration channels are in units of ppm where the full scale is defined in the
software and in the configuration of the Moxa ioLogik E1240 Modbus device.
o Full scale is configured in the software in the Setup menu under the 4–20 mA tab (see Figure
28), where the full scale in ppm is defined for 20 mA output (or full scale of Moxa).
o For registers 9 – 24 to report the correct gas concentration, the Moxa must be configured to
have the same full scale as in the software. (The factory default scale must be set to the same
value using the included Moxa ioSearch software.)
Table 7 Moxa Registers, Type, Corresponding 4-20 mA Channel, and Channel Name
Register Type
4 – 20 mA Channel
Channel Name
Temperature Warning/Alarm
Scaling Value Hi (float LSBF)
5 Temperature Warning/Alarm
Scaling Value Hi (float LSBF)
6 Analyzer Warning/Alarm
Scaling Value Hi (float LSBF)
1 Gas 1 Concentration
Scaling Value Hi (float LSBF)
2 Gas 2 Concentration
Scaling Value Hi (float LSBF)
3 Gas 3 Concentration
4 Gas 4 Concentration
USB Drive – Local Data Access
The ABB GLA531 Series LGR-ICOS™ Gas Analyzer supports up to USB 2.0. Refer to the File Transfer section
(also, see Figure 22) for instructions on transferring data from the analyzer to a USB 2.0 memory stick.