GLA531 Series LGR-ICOS™ Gas Analyzer
Operating Manual
3KXG165002R4201_2021_04
ABB PROPRIETARY INFORMATION 71
Table 19 Possible Failed Components Based on Functions
Function Impact Components
System start-up
(or computer boot-up)
Analyzer does not start
Analyzer crashed and shut down
PC104 Computer Stack
Hard Disk Drive
DC-to-DC Converter
Ethernet or USB
communications
No direct access to computer
No I/O access
PC104 Computer Stack
Ethernet cable or USB cable
Remote digital I/O access
No connections to Modbus
ioLogik E1240 Modbus unit
Ethernet cable installation
4-20 mA analog signal
transmissions
No analog data transmissions
4-20 mA Isolator unit
Signal cable installation
Temperature control of
analyzer
Analyzer alarm or warning
Measurement drift or error
Condensation inside ICOS cell
Heater controller
Heater element
Thermocouple (T/C)
Solid state relay
Gas Temperature Reading
(on HMI Control Bar)
Measurement drift or error
Display wrong temperature
ICOS TH-Amp heat sensor
assembly
Pressure control of analyzer
Analyzer alarm or warning
Measurement drift or error
Abnormal analyzer spectrum
profile
Exhaust pump
Solid state relay
2-Port Solenoid Driver
PWM pressure controller
EPC valve
Inlet solenoid valve
ICOS Pressure Gauge
ICOS Inlet Orifice
ICOS Gas filters
Light intensity detection
(low or zero detector voltage
on spectrum screen)
Analyzer alarm
ICOS function failure
No measurement outputs
Laser
Laser driver board
Photo detector
Mirror reflectivity
(decreased laser ring-down
time readings on HMI Control
Bar)
Analyzer alarm or warning
Decrease in ring-down time
Measurement error
Dirty astigmatic mirrors
Purge System
If the purge alarm triggers for X-purge systems, and power to the analyzer is shut off:
Check that the enclosure door is completely closed and all of the locking clamps are engaged.
Check that the purge-gas supply pressure is adequate and regulated to the specified purged-air
regulated pressure on the certification plate.