EasyManua.ls Logo

ABB LGR-ICOS GLA531 Series - Controllers and Relays

ABB LGR-ICOS GLA531 Series
71 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Manual
3KXG165002R4201_2021_04
ABB PROPRIETARY INFORMATION 57
GLA531 Series LGR-ICOS™ Gas Analyzer
Figure 45 DC-to-DC Voltage Converter Board
Temperature Controller
Depending on the selected options, up to four temperature controllers are inside the analyzer (see Figure 46
below). Each temperature controller receives a K-Type thermocouple (T/C) input. It has a single voltage output to
control a DC solid-state relay as shown in Figure 47.
Figure 46 Temperature Controller
DC Solid-State Relay
DC solid-state relays (SSR) inside the ICOS Module switch the heaters on and off, and turn on the exhaust
pump. The SSR for the heaters receive their enable/disable commands from the temperature controllers. The
SSR for the exhaust pump receives the control signal from the two-port solenoid driver on channel 2. When an
SSR is enabled (energized), the green LED on top illuminates.