GLA531 Series LGR-ICOS™ Gas Analyzer
Table 9 Channel 6 Analyzer Warnings/Alarms
Warning /
Alarm
Current
(mA)
UI Display
4-20 mA Alarm Status
Detected Problem
Alarm 4 ± 0.1 Data Health (A)/(B) Laser A and/or B goodness of fit is poor
Alarm 5 ± 0.1 Pressure Pressure is not in operating range
Alarm 6 ± 0.1 HD Space
Hard drive (HD) space is low, deleting oldest
files
Alarm 7 ± 0.1 Mirror Health (A)/(B) Mirror health has degraded, clean mirrors
Alarm 8 ± 0.1 Linelock (A)/(B)
Laser A and/or B peak position is outside of
control range, contact customer support
Alarm 9 ± 0.1 Signal Power (A)/(B)
Laser A and/or B power has degraded, contact
customer support
Alarm 10 ± 0.1 Maintenance Maintenance is needed on system now
Warning 12 ± 0.1 Data Health (A)/(B) Laser A and/or B goodness of fit is not optimal
Warning 13 ± 0.1 Pressure Pressure is noisy
Warning 14 ± 0.1 HD Space HD space is low
Warning 15 ± 0.1 Mirror Health (A)/(B) Mirror health is degrading, clean mirrors soon
Warning 16 ± 0.1 Linelock (A)/(B)
Laser A and/or B peak position is moving very
fast
Warning 17 ± 0.1 Signal Power (A)/(B)
Laser A and/or B power is degrading, contact
customer support soon
Warning 18 ± 0.1 Maintenance
Maintenance is needed on system soon
No issue 20 ± 0.1 No warning/alarm
NOTE
deactivated for more than 10 minutes, allow the
heaters to reach the set temperature before CH5 and CH6 start porting 20
mA values.
An analyzer not equipped with a gas-expansion heater, orifice heater, or
inlet-tube heater, does not have the associated color flags in the Alarm
Status display screen.