25
Figure 25  HMI Display Keyboard
The keys (1), (4), (2) and (3) are available for the menu-controlled 
configuration.
•  The menu / sub-menu name is displayed above in the HMI 
display.
•  The number / line of the currently selected menu item is 
displayed in the upper right of the HMI display.
•  A scroll bar is located on the right edge of the HMI display and 
shows the relative position of the currently selected menu 
item within the menu.
•  Both of the keys (1) and (4) can have various functions. The 
meaning of these buttons is displayed below in the HMI 
display above the corresponding button.
•  The user can browse through the menu or select a number 
within a parameter value using both keys (2) and (3). The 
button (4) selects the preferred menu item.
Commissioning using the Easy Setup Menu
The most common configuration parameters are summarized 
in the Easy Setup Menu. This menu provides the quickest way to 
configure the device.
For a detailed description of all of the device menus and 
parameters, see the Operation section of this manual.
1  Log on to the LMT at the Standard or Advanced access level.
2  Select Easy Setup in the main menu
3  Select a language in the “Easy Setup” menu and press 
4  Select available option of primary variable in the “Easy 
Setup” menu and press 
5  Select available option of units of measure in the “Easy 
Setup” menu and press 
6  Set LRV in the “Easy Setup” menu and press 
LRV is the lower range value which corresponds to 4 mA current 
out value
7  Set URV in the “Easy Setup” menu and press 
URV is the upper range value which corresponds to the 20 mA
current out value
8  Set Damping time in the “Easy Setup” menu and press
Damping allows smoothing step response in the device output.