Section 7 Module Replacement Detector Module Replacement
AA003700-01 rev. G. 2.0 29
Section 7 Module Replacement
Detector Module Replacement
Removing the Detector Module
STEP 1 With a 4 mm Hex driver, unscrew the 3 captive screws holding the Detector
module into place.
STEP 2 Remove the Detector module from the spectrometer: The module is tightly
fitted into the housing, gently pull the module to remove it from the
spectrometer.
Screws are easily inserted and tighten. Do not use excessive force to set the
screws in place, otherwise spectrometer could be damaged.
The type of Detector module being installed into the spectrometer, is
automatically detected.
The Detector module can be removed and installed without powering OFF the
instrument.
Figure 7- 1. Detector module