Removing the Source Module Section 7 Module Replacement
32 AA003700-01 rev. G. 2.0
STEP 2 Remove with a 2.5 mm Hex driver the 4 screws holding the front panel in
place.
STEP 3 Slide down and remove the front panel from the spectrometer.
STEP 4 Unscrew with a 2.5 mm Hex driver the bracket holding the Source Electronic
Board.
Figure 7- 3. Front panel captive screws
Do not discard the four screws: they will be required to reinstall the front panel.
Figure 7- 4. Source module: Source control board and source