FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D 69
10 Specification
Note
The device data sheet is available in the ABB download area at
www.abb.com/flow.
Permitted pipe vibration
In accordance with EN 60068-2-6
Valid for sensors in remote mount and integral mount design.
Maximum deflection: 0.15 mm (0.006 in) in the frequency
range of 10 to 58 Hz
Maximum acceleration: 2 g in the frequency range of
58 to 150 Hz
ProcessMaster – Temperature data
The temperature range offered by the device is dependent on a
number of different factors.
These factors include the measuring medium temperature
T
medium
, the ambient temperature T
amb
, operating pressure
P
medium
, liner material and the approval for explosion protection.
Storage temperature range
Æ40 to 70 °C (Æ40 to 158 °F)
Maximum permissible cleaning temperature
CIP media Liner Cleaning temperature
Steam PTFE, PFA 150 °C (302 °F)
Cleaning fluid PTFE, PFA 140 °C (284 °F)
The maximum cleaning temperature specified applies to a
maximum ambient temperature of 25 °C (77 °F).
If the ambient temperature up-scales > 25 °C (> 77 °F), then
the temperature difference to the current temperature must
be subtracted from the max. cleaning temperature.
The specified cleaning temperature may have an effect for a
maximum of 60 minutes.