116  FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D   
 
… 9  Diagnosis / error messages 
… Overview 
 
Error no. / Range  Error text  Current output  Digital output  Pulse output  LCD display  Error maskable? 
S065.044 / 
operation 
Inhouse temp. alarm. 
Reduce ambient temperature. 
Current value - no 
change. 
No Answer  No Answer  Current value - no 
change. 
 
S064.041 / 
operation 
EPD alarm. 
Secure pipe is completely filled. 
Alarm - configured 
as in menu ‘Curr. at 
EPD Alarm’. 
Alarm, if DO as 
‘Logic / Alarm Signal
/ EPD Alarm’ is 
configured 
0 Hz  0 %  Menu ‘...Group 
Masking’. 
S063.040 / 
operation 
TFE alarm. 
Secure pipe is completely filled. 
Alarm - configured 
as in menu ‘Curr. at 
TFE Alarm’. 
Alarm, if DO as 
‘Logic / Alarm Signal
/ TFE Alarm’ is 
configured 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S062.039 / 
operation 
Sensor temp. limits alarm. 
Change limits or change fluid 
temperature. 
 
Current value - no 
change. 
No Answer  Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S061.038 / 
operation 
Conductivity limits alarm. 
Change limits or Check 
application. 
Current value - no 
change. 
Alarm, if DO as 
‘Logic / Alarm Signal
/ Conductivity’ is 
configured 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S060.037 / 
operation 
Gas bubble alarm. 
Check conditions of application. 
Current value - no 
change. 
Alarm, if DO as 
‘Logic / Alarm Signal
/ Gas bubble Alarm’ 
is configured 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S052.016 / 
operation 
Curr.Out 31 / 32 is saturated. 
CO process value out of range. 
Adapt Qmax. 
Alarm - configured 
as in menu 
‘„Curr.Out > 20.5mA’.
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S051.017 / 
operation 
Curr.Out V1 / V2, V3 / V4 
saturated. 
CO process value out of range. 
Adapt Qmax. 
S049.019 / 
electronics 
Option Card 1 com error. 
Defective Card. Check Card 1. 
Call Service. 
Does not react 
anymore 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S048.020 / 
electronics 
Option Card 2 com error. 
Defective Card. Check Card 2. 
Call Service. 
Does not react 
anymore 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’.