FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  117 
 
Error no. / Range  Error text  Current output  Digital output  Pulse output  LCD display  Error maskable? 
S047.015 / 
operation 
Pulse output is cut off. 
Wrong config. Check pulse out 
configuration. 
Current value - no 
change. 
Current value - no 
change. 
Largest possible 
pulse rate 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S046.000 / 
operation 
Mass flowrate exceeds limits. 
Check flowrate and alarm limits. 
Current value - no 
change. 
Alarm, if DO as ‘Qm 
Massflow Max’ Or 
‘Qm Massflow Min’ 
is configured. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S044.001 / 
operation 
Volume flowrate exceeds limits. 
Check flowrate and alarm limits. 
Current value - no 
change. 
Alarm, if DO as ‘Qv 
Volumeflow Max’ Or 
‘Qv Volumeflow Min’ 
is configured. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
S041.034 / 
electronics 
DC feedback  regulation. 
Check conditions of application. 
Call service. 
4 mA (0 % flow)  No Answer.  0 Hz  0 %  Menu ‘...Group 
Masking’. 
S040.031 / 
electronics 
Coil Inductance alarm. 
Call service. 
Current value - no 
change. 
No Answer.  Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
M038.009 / 
electronics 
Sensor memory defective. 
Mem. or connect. defective. 
Replace memory. 
Current value - no 
change. 
No Answer.  Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
M037.014 / 
Electronics 
NV chips defect on 
Motherboard. 
Defective MB. Replace MB. Call 
Service. 
M032.022 / Config.
Curr.Out 31 / 32 not calibrated. 
Call Service. 
M031.023 / Config.
Curr.Out V1 / V2 not calibrated. 
Replace Current Option Card. 
Call Service. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
M030.024 / 
Config. 
Curr.Out 31 / 32 not calibrated. 
Call Service. 
M028.007 / 
Config. 
Display value is < 1600 h at 
Qmax. 
Change mass Unit or vol. Unit 
for Totalizer. 
Current value - no 
change. 
No Answer.  Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
M026.004 / 
Config. 
Maintenance interval is reached. 
Perform maintenance. 
M024.008 / 
Config. 
Device not  calibrated. 
Call Service. 
High Alarm or Low 
Alarm, depending on
parameter ‘Curr.Out 
at Alarm’. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu '...Group 
Masking’. 
M020.027 /  
Electronics 
Communicat. Card not 
responding 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu ‘...Group 
Masking’. 
 
 
Change from one to tw o columns