10 CI/FSV/FSS/430/450-EN Rev. G | VortexMaster FSV430, FSV450 SwirlMaster FSS430, FSS450
2.3.3 Temperature data
Operating temperature ranges:
— The ambient temperature range T
amb.
is -40 ... 85 °C
(-40 ... 185 °F). This is dependent on the temperature
class and measuring medium temperature, as listed in the
following tables.
— The measuring medium temperature T
medium
is -200 ... 400 °C (-328 ... 752 °F).
Devices without LCD indicator and with HART
communication
Temperature class T
amb.
max. T
medium
max.
T4 ≤ 85 °C 90 °C
≤ 82 °C 180 °C
≤ 81 °C 280 °C
≤ 79 °C 400 °C
T4 ≤ 70 °C 90 °C
≤ 67 °C 180 °C
≤ 66 °C 280 °C
≤ 64 °C 400 °C
T5 ≤ 56 °C 90 °C
≤ 53 °C 180 °C
≤ 52 °C 280 °C
≤ 50 °C 400 °C
T6 ≤ 44 °C 90 °C
≤ 41 °C 180 °C
≤ 40 °C 280 °C
≤ 38 °C 400 °C
Devices without LCD indicator and with Modbus
communication
Temperature class T
amb.
max. T
medium
max.
T4 ≤ 85 °C 90 °C
≤ 82 °C 180 °C
≤ 81 °C 280 °C
≤ 79 °C 400 °C
T4 ≤ 70 °C 90 °C
≤ 67 °C 180 °C
≤ 66 °C 280 °C
≤ 64 °C 400 °C
T5 ≤ 40 °C 90 °C
≤ 37 °C 180 °C
≤ 36 °C 280 °C
≤ 34 °C 400 °C
T6 ≤ 40 °C 90 °C
≤ 37 °C 180 °C
≤ 36 °C 280 °C
≤ 34 °C 400 °C
Devices with LCD indicator, order code L1
Temperature class T
amb.
max. T
medium
max.
T4 ≤ 85 °C 90 °C
≤ 82 °C 180 °C
≤ 81 °C 280 °C
≤ 79 °C 400 °C
T4 ≤ 70 °C 90 °C
≤ 67 °C 180 °C
≤ 66 °C 280 °C
≤ 64 °C 400 °C
T5 ≤ 40 °C 90 °C
≤ 37 °C 180 °C
≤ 36 °C 280 °C
≤ 34 °C 400 °C
T6 ≤ 40 °C 90 °C
≤ 37 °C 180 °C
≤ 36 °C 280 °C
≤ 34 °C 400 °C