CAUTION:
▪ The device complies with the limits for medical devices contained in IEC 60601-1-2:2014-04,▪
CISPR11:2015 (BS EN 55011:2016), and ETSI EN 300 328 V2.1.1 (2016‑11). However, the device may
cause radio interference or may disrupt the operaon of nearby equipment. It may be necessary to
migate this eect by reorienng or relocang the receiving device.
▪ To avoid increasing emissions or decreasing immunity from a device or system, use only components▪
approved by Abbo Medical with this system. Do not use Abbo Medical components with devices or
systems that are not approved by Abbo Medical.
Table 7. Guidance and Manufacturer’s Declaraon – Electromagnec Emissions
Emissions Test Compliance Electromagnec Environment Guidance
RF emissions
CISPR 11
Group 1 The device produces RF emissions as a result of its internal and system interface funcons. Its
RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all establishments including domesc establishments and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuaons/ icker
emissions
IEC 61000-3-3
Not applicable
41
▪
▪