Aer the device has been implanted and the pocket is closed, interrogate the device and set the Lead Type to
the correct seng. Lead Type sengs are described on the programmer's on‑screen help. For MR Condional
pulse generators, set the Addional Hardware parameter to the correct seng.
NOTE:
▪ In CRT‑Ps, the right and le ventricular pulse amplitudes and pulse widths are independently▪
programmable. The pulse amplitude and pulse width should be evaluated in each chamber accordingly.
▪ In CRT‑Ps, independent lead impedance values are displayed for the RV and LV leads.▪
Managing and Following Paents
Paent Registraon Form
Fill out and return the Paent Registraon Form to register the paent and facilitate paent tracking.
Paent Idencaon Card
Fill out the Paent Idencaon Card as provided in the outer box with the paent’s name, implant date, and
your name or healthcare facility informaon along with the contact phone number. Record device informaon
(model number, serial number, and UDI number) and leads informaon (model number, serial number, and
UDI number), or apply the labels provided with the sterile package in the designated secons of the card. Fill
out the locaon of the device in the corresponding implant locaon eld on the Paent Idencaon Card.
Fill out the model/length and implant locaon of the leads in the corresponding elds on the Paent
Idencaon Card.
Paent Follow‑Up
Give the completed card to the paent. To obtain a replacement card if a paent loses or damages their card,
contact
Abbo
Medical
Technical Support (page 38).
34