Call Customer Care Toll Free at 1-888-522-5226 www.AbbottDiabetesCare.com
ENGLISH
Ensuring Accurate Control Solution Results
Use only FreeStyle® Control Solution with your FreeStyle Freedom™
System.
Replace the cap on the control solution bottle immediately after using.
Do not use control solution past the expiration date, this may cause
inaccurate results.
Do
not add water or any liquid to control solution.
Control solution tests are specified to be accurate only when
performed between 59º to 104º F
(15º to 40º C).
The control solution test results should fall within the range printed
on the test strip vial label.
Important: The control solution range is a target range for control
solution test results only. It is not a target range for your blood glucose
level or blood glucose test results.
If control solution results are out of this range,
repeat the test.
If your control solution test results continue to fall out of range,
do
not use the FreeStyle Freedom System to test your blood glucose.
Contact Customer Care at 1-888-522-5226.
Caution: Results from FreeStyle Control Solution tests do not
reflect your blood glucose level.