14
Call Customer Care Toll Free at 1-888-522-5226 www.AbbottDiabetesCare.com
ENGLISH
The result will appear on the display screen and if the sound is
on, your meter will beep twice. To mark the result as a control
solution test result, press and hold the
cbutton for two seconds
until the control solution symbol (
) appears at the top of the
display screen.
I
f you have marked a result as a control solution test by mistake,
you can unmark it by pressing and holding the
cbutton again
for two seconds.
Compare your control solution test result to
the range printed on your test strip vial. Your
control solution test results should fall within
this range. The range shown in picture is only
an example.
4
Out of range test results may be caused by:
Expired or bad control solution
Expired or bad test strip
Error in performing test
Watered-down control solution
Code on test strip vial does not match code set in meter
Meter malfunction
Control solution test done outside 59º to 104º F
(15º to 40º C)
To power off the meter, remove the test strip.